Vacuum Viscometers - Senze-Instruments Benelux

Senze-Instruments BENELUX
Go to content
Process & Quality Control

Vacuum Viscometers

Coatings & Industry > Physical properties > Viscosity & Consistency > Glass Capillary Viscometers
Vacuum Viscometers
In most glass capillary  viscometers, the  samples flow  under gravity.
When liquids are too viscous to flow  readily under gravity, vacuum viscometers may be  used to measure viscosity (in mPa•s or cP).
In these instruments a vacuum is applied to one end of the  viscometer to pull the  liquid through the  capillary  into the  timing bulb(s).
CANNON offers several  types of vacuum viscometers, including the  Cannon-Manning Vacuum , the  Asphalt Institute Vacuum, and the  Modified Koppers Vacuum.
Like the  Cannon-Fenske Opaque viscometer, these are all “reverse  flow” viscometers.
Vacuum viscometers require a vacuum that is very  accurately controlled. The CANNON
DVR-1000 or DVR-1500 Digital Vacuum Regulator is an ideal instrument for maintaining constant vacuum.



CANNON-MANNING VACUUM VISCOMETERS

Cannon-Manning Vacuum
Calibrated CMVC Series
For measurement of viscosity of highly viscous materials such as asphalt  cement at 60°C (140°F) according to ASTM D 2171.
Applicable to materials with viscosity range of 0.036 to 80 000 poise.
Vacuum of 300 mm Hg is applied to small arm with timing bulbs; two bulbs are incorporated to extend viscosity range of viscometer.
  • Minimum sample volume 6 mL. Require liquid bath depth of 180 mm.
  • Provided with certificate of calibration and instruction sheet.
  • Holder  is not supplied. For holder, see 9726-N15.

Cannon-Manning Vacuum Calibrated CMVC-RO Series
Same as 9724-H50 series, but with permanently attached round metal holder, for hole 51 mm in diameter.

ASPHALT INSTITUTE VACUUM VISCOMETERS


Asphalt Institute Vacuum Calibrated AIVC Series
Similar to 9724-H50 series, but with graduated capillary  instead of two timing bulbs.
  • Minimum sample size 3 mL. Require bath depth of 180 mm.
  • Note: For sizes 400R and 800R, the “roofing” design is normally used for segment B, below, which extends the viscosity range to 1 400 000 and 5 800 000 poise, respectively. See ASTM D 2171, Table X2.
  • Provided with certificate of calibration and instruction sheet.
  • Holder  is not supplied. For holder, see 9726-N15.


Asphalt Institute Vacuum
Calibrated AIVC-RO Series Same as 9724-R50 series, but with permanently attached round metal holder,
for hole 51 mm in diameter.


CANNON MODIFIED KOPPERS VACUUM VISCOMETERS
Calibrated MKVC Serie-s
For measurement of viscosity of asphalts  at 60°C in accordance with ASTM Standard D2171.
Applicable to materials with viscosity range of 42 to 200 000 poise.
Two piece design, with sample reservoir  and calibrated capillary. Vacuum of 300 mm Hg must be  applied to top  of capillary tube.
  • Minimum sample volume 2 mL. Require liquid bath depth of 178 mm.
  • Provided with certificate of calibration and instruction sheet.
  • Holder  is not supplied. For holder, see 9726-N30.


REMOVABLE HOLDERS FOR CANNON VISCOMETERS
All CANNON viscometer holders fit holes 51 mm (2-inches) in diameter.
Rubber holders are made of neoprene rubber.  
Plastic holders hold the viscometer firmly by means of a spring clip.
Metal holders are constructed of chrome-plated machined brass and are provided with a handle.
                                                                                       Type of Holder
Viscometer                                                       Plastic  Rubber  Metal

Cannon-Manning Vacuum                        —              H140        —
Asphalt Institute Vacuum                          —              H140        —
Modified Koppers Vacuum                       —              H141       

Catalog Numbers For Viscometer Holders
Holder        Catalog #  

H140            9726-N15
H141            9726-N30