Datalogger ECOLOG TH1
Connection for 1 integrated (-35°C ..55°C; 0 ..100%rH) or up to 2 external humidity and temperature sensors (-35°C ..55°C; 0..100%rH) or a second temperature sensor (-50°C ..140°C)
Calibrated, interchangeable humidity and temperature sensors
Memory capacity for 64 000 data points
Large display with alarm information
4 button keypad, External Start, Alarm Reset, Measurement and Alarm Scroll
1 digital input
DB15 multifunction connector with alarm output and RS232
Direct connection to printer for data and alarm printout  

Alarm accessories available : Flasher lamp, Flasher lamp with siren and flash light, Automatic telephone dialing unit


Datalogger ECOLOG TH2
For 2 external, calibrated and interchangeable humidity and temperature sensors -35°C ..70°C;0% .. 100% rH  
Memory capacity for 64 000 data pointsoLarge display with alarm information
4 button keypad, External Start, Alarm Reset, Measurement and Alarm Scroll
1 digital input
DB15 multifunction connector with alarm output and RS232
Direct connection to printer for data and alarm printout
Alarm accessories available :  Flasher lamp, Flasher lamp with siren and flash light, Automatic telephone dialing unit



Different configuration are available :


More configurations here.


Just clic under that link, and you will discover a complete report sample.



With the Elprolog Analyse software this report is printed in two clics.








Holding plate for THx
Part No.: 102-2805-CR
with 2 DB15 connectors for external probes

Terminal screws for 1 digital input
Alarm
Stainless steeland DB15 / RS232 connection


Accessories for the object dewpoint calculation :

NTC Surface temperature probe
Part No.: 102-3041-L03
Temperature range : -50°..+140°C
Sensor flat surface area 6 x 30 mm
Can be screwed down
Silicone cable
L=3m
Aluminium


Protection cage for ECOLOG TH1, stainless steel. Ref 102-2804-F



Other accessories under :  http://www.elpro.com


Protective roof for ECOLOG TH1With fixation bracket lockableStainless steel, no weather protectionOnly for splash-water protection and drip proof indoor applications.  Ref 102-2804 G

2) Periodical recalibration:
With SCS calibration solutions and calibration device by the end user or by Senze-Instruments service center.
Alternatively there is the possibility to get a calibrated sensor as an interchangeable part from Senze
-Instruments.
Interval: every 12 months in clean operating environment; in environment with high humidity, dust, smoke etc, every 6 months or in case of doubt.
Required for humidity calibration and adjustment:
Extension cable for ECOLOG THx for calibration: Part No.: 102-3215-S01
Calibration unit for humidity logger: Part No.: 102-2902
Calibration ampullae (set of 5) Part No.: 102-2901-Hx, x= 0%, 35%, 50%, 80%, 95%
NB: SCS = SWISS CALIBRATION SERVICE

Check / Verification of the Relative Humidity Measurement
1) New devices :
All dataloggers are factory-calibrated with SCS calibration solutions and subsequently receive a calibration certificate. The adjustment points are 0%rH and 80%rH (95%rH for high levels of humidity). The humidity sensors are calibrated and interchangeable.
The calibration values are read in by the logger.